Operating instructions
The menu
22
tESt
fcESt
CRL
m
E
off
fc \
\
\
Test of the balance with internal weight
In this setting the accuracy test of the balance is performed with the internal
weight.
Test of the balance with external weights
The accuracy of the balance can be checked with any external weight.
You will find information on how to perform the adjustment and test function in
Sections 2.5, 5.6, 5.7 and 5.8.
Switching off the adjustment and test function
The adjustment or test is switched off with the CAL key.
Note
Irrespective of the settings in the menu option Adjustment (Calibration), the fully
automatic adjustment FACT is always active with certified balances.
4.5 Switching automatic adjustment call-up on or off
In this menu option you can switch the call-up of the automatic adjustment (calibration) or test on or off.
Note: If you have set «FACT» in the menu option Adjustment (calibration), the automatic adjustment call-up is always active
and will thus be skipped in the menu. It becomes active again as soon as «FACT» is switched off.
The following settings are available:
Automatic adjustment (calibration) or test call-up switched on
This is the factory setting. The balance prompts you with a flashing «Cal» in the
display to adjust (calibrate) or test it with the internal weight or external weights.
The call-up is triggered by, e.g. temperature changes of the surroundings.
Cal
• nfo on
Cal
Info off
Automatic adjustment (calibration) or test call-up switched off
The automatic adjustment or test call-up is switched off.
Note
With the certified versions of balances, cal I up of the automatic adjustment or test
can not be selected, i.e. FACT is always active.