Operating instructions
The menu
2?
4.3 Reset
In this menu option you have the possibility to reset all menu settings to the factory setting.
long
•ESEfc
donE
Resetting settings to factory setting
If you select this option and then save and quit the menu, all menu settings are
reset to the values set in the factory.
Before the return to the weighing mode, the resetting is briefly confirmed in the
display.
4.4 Selection of the adjustment (calibration) and test function
Your balance can be adjusted (calibrated) with internal or external weights. The balance can also be checked by a test
with internal or external weights. If you have attached a printer to your balance, the data of the adjustment (calibration)
and the results of the test are printed out following GLP recommendations.
The following settings are available:
FRCt
CRL .nfc
Fully automatic internal adjustment (calibration) FACT (Fully
Automatic Calibration Technology)
This is the factory setting. The balance adjusts (calibrates) itself fully automat-
ically
- after the warm-up phase following connection to the power supply,
- when a change in the ambient conditions, e.g. the temperature could lead to
a noticeable measurement deviation,
- with certified balances, always independent of the settings in menu option
Adjustment.
Internal adjustment (calibration)
Adjustment (calibration) is performed at a keystroke with the built-in weight.
<j i ir i (_'•(_
Adjustment (calibration) with external weights (VariCal)
Adjustment (calibration) is performed with a selectable* external weight.
* This function is blocked for certified balances.