Operating instructions

Model ML1602 ML3002 ML3002E
Minimum weight (acc. to USP) 30 g 30 g 30 g
Minimum weight (U=1%, k=2) 2 g 2 g 2 g
Minimum weight (OIML) 0.5 g 0.5 g 0.5 g
Settling time, typ. 1.5 s 1.5 s 1.5 s
Weighing technology MonoBloc MonoBloc MonoBloc
Weighing pan dimensions (WxD) [mm] 170x190 170x190 170x190
Balance dimensions (WxDxH) [mm] 184x290x84 184x290x84 184x290x84
Net Weight [kg] 3.6 3.6 3.4
Model ML4002 ML4002E
Maximum load 4200 g 4200 g
Readability 0.01 g 0.01 g
Taring range 0...4200 g 0...4200 g
Repeatability (sd) 0.01 g 0.01 g
Linearity 0.02 g 0.02 g
Sensitivity temperature drift (10...30°C) 3 ppm/°C 3 ppm/°C
Internal adjustment yes no
Adjustment range with external weights 2000...4200 g 2000...4200 g
Weights for routine testing
Large Weight/Class OIML/ASTM
Small Weight/Class OIML/ASTM
2000 g / F2/4
200 g / F2/4
2000 g / F2/4
200 g / F2/4
Minimum weight (acc. to USP) 30 g 30 g
Minimum weight (U=1%, k=2) 2 g 2 g
Minimum weight (OIML) 0.5 g 0.5 g
Settling time, typ. 1.2 s 1.2 s
Weighing technology MonoBloc MonoBloc
Weighing pan dimensions (WxD) [mm] 170x190 170x190
Balance dimensions (WxDxH) [mm] 184x290x84 184x290x84
Net Weight [kg] 3.6 3.4
Model ML6002 ML6002E
Maximum load 6200 g 6200 g
Readability 0.01 g 0.01 g
Taring range 0...6200 g 0...6200 g
Repeatability (sd) 0.01 g 0.01 g
Linearity 0.02 g 0.02 g
Sensitivity temperature drift (10...30°C) 3 ppm/°C 3 ppm/°C
Internal adjustment yes no
Adjustment range with external weights 2000...6200 g 2000...6200 g
Weights for routine testing
Large Weight/Class OIML/ASTM
Small Weight/Class OIML/ASTM
5000 g / F2/4
200 g / F2/4
5000 g / F2/4
200 g / F2/4
Minimum weight (acc. to USP) 30 g 30 g
Minimum weight (U=1%, k=2) 2 g 2 g
Minimum weight (OIML) 0.5 g 0.5 g
Settling time, typ. 1.2 s 1.2 s
Weighing technology MonoBloc MonoBloc
Weighing pan dimensions (WxD) [mm] 170x190 170x190
Balance dimensions (WxDxH) [mm] 184x290x84 184x290x84
Net Weight [kg] 3.6 3.4
85Technical Data