Operating instructions
Sample Printout:
--- Statistical Info ---
21.Jan 2009 11:34
METTLER TOLEDO 
Balance Type ML4002
SNR 1234567890
SW V1.00
------------------------
Operating time 18d 23h
Total weight loaded
 115.7191 kg
Number of weighings
 1255
Number of key presses
 4931
Motor movements
 1012
Backlight operating time
 18d 4h
Next service due date
 01.01.2010
------------------------ 
18.6 Calibration History
The "Calibration History" function allows you to view and print information of the last 30 (thirty) balance adjust
ment. Adjustments made by a service technician and normal user are counted together.
a)  Press «
» to start "CAL.HIST".
b)  Press « » for printout.
c)  Press « » key to scroll forward through the displayed list of Adjustments history information.
d)  Press «C» to cancel the test procedure. The balance will return to the topic "DIAGNOS.".
Sample Information Displayed:
Note Display
S = External adjusted service 05:03:09S  01
-3 PPM
I = Internal adjusted 04:03:09I 02
2 PPM
·
·
·
·
·
·
I = Internal adjusted 03:03:09I 28
-1 PPM
E = External adjusted user 02:03:09E 29
4 PPM
I = Internal adjusted 02:03:09I 30
1 PPM
70 Application "Diagnostics"