Operating instructions
18.4 Motor Test
The motor test allows you to test the calibration motor of the balance.
Note: On models with internal weight only.
a) Press « » to start "CAL.MOT.T".
"RUNNING" is displayed during the Motor Test. A motor test is deemed successful when all the motor posi
tions have been successfully tested. At the end of the test, the test information will be printed.
b) Press « » for printout.
c) Press «C» to cancel the test procedure. The balance will return to the topic "DIAGNOS.".
Sample Printout:
------ Motor Test ------
21.Jan 2009 11:34
METTLER TOLEDO
Balance Type ML204
SNR 1234567890
SW V1.00
Motor Test OK
------------------------
18.5 Balance History
The balance history function allows you to view and print the history of the balance.
a) Press « » to start "BAL.HIST" .
b) Press « » to scroll forward through the displayed list of balance history information.
c) Press «
» for printout.
d) Press «C» to cancel the test procedure. The balance will return to the topic "DIAGNOS.".
Sample Information Displayed:
Information Display
Operation Time (year:day:hour) 00:018:04
Total load kg (t) 115.7191 kg
Number of weighings 1255
Number of key pressed 4931
Number of motor movements 1012
Backlight time (year:day:hour) 00:018:04
Next service due date 01:01:10
69Application "Diagnostics"