Sample Printout:
-- Repeatability Test --
21.Jan 2009 11:34
METTLER TOLEDO
Balance Type ML4002
SNR 1234567890
SW V1.00
Temperature 21.3 °C
No. of tests 10
------------------------
1. Time 00:00:00
1. Temp 21.3 °C
2. Time 00:00:04
2. Temp 21.3 °C
.
.
.
------------------------
s Dev 0.004 g
Max Temp 21.3 °C
Min Temp 21.3 °C
Mean Temp 21.3 °C
Total Time 00:00:44
------------------------
Examples:
Repeatability test is a tool to do functional check with the balance. It may be performed:
To check function of balance
during installation to store print out with installation documents.
after preventative maintenance to store print out with installation maintenance report.
when remarkable decrease of weighing performance occurs, so that you can email/fax print out to ser
vice support provider for diagnose purposes.
To develop the optimal environment settings (see menu topic "ENVIRON.").
Measure the time you need to perform repeatability test with each "STABLE", "STD." and "UNSTAB." setting.
The setting with the fastest total time suits best for the existing environmental conditions.
18.2 Display Test
The display test allows you to test the display of the balance.
a) Press «
» to start "DISPLAY".
All possible segments and icons on the display will illuminate.
b) Press «
» to print the test information.
c) Press «C» to cancel the test procedure. The balance will return to the topic "DIAGNOS.".
67Application "Diagnostics"