Operating instructions
LMT
WARN.
1.25
g
Setting the Warning Limit
The default value of the warning limit:
Warning limit = control limit / safety factor
Example: 2.5 g / 2 = 1.25 g.
a) For changing the value, press «+» to scroll up or «–» to scroll
down. Progressing speed by press and hold.
b) Press « » to confirm the value.
Note: The default values of control limit and the warning limit are evaluated according the GWP recommenda
tion. These are based under the assumption that the weighing process tolerance is 0.1% and the safety factor
is 2.
PRESS EN
PRESS EN
5000.00
g
On completion of the setting procedure, your balance is ready for
the routine test procedure.
Note: The test weight must be acclimatized to the ambient tempera
ture of the balance.
a) Press « » to start the test.
b) Follow the instructions on the display. If the test weight value is
flashing: Load the test weight (displayed value).
The printout starts after the weighing pan is unloaded.
Exit the current test procedure:
Press and hold « » for executing a new application.
Printout:
----- Routine Test -----
21.Jan 2009 12:56
METTLER TOLEDO
Balance Type ML6001
SNR 1234567890
Sensitivity:
Test weight 5000.00 g
Value 5000.11 g
Warning L. 1.25 g
Control L. 2.50 g
Warning L. OK
Control L. OK
Signature
........................
------------------------
What if Warning Limit or Control Limit are "FAILED"?
The "SOP for Periodic Sensitivity Tests (Routine Tests)" provides information about measures when routine tests
fail. Find a download version of these SOPs on www.mt.com/gwp, link "GWP
®
The Program / Routine Opera
tion".
64
Application "Routine Test"