Operating instructions

20.0
21.8
If you have selected water or ethanol as the auxiliary liquid:
a) Enter the current temperature of the auxiliary liquid (read off on
thermometer). Change the value by scrolling up «+» or down «-».
The temperature ranges from 10 °C to 30.9 °C.
b) Press « » to confirm the value.
Note: The densities of distilled water and ethanol in the range 10 °C to
30.9 °C are stored in the balance.
01.00200
01.00000
If you have selected a freely definable auxiliary liquid:
Enter the density of the auxiliary liquid at the current temperature (read
off on thermometer).
a) Press « » to select a digit. The selected digit is blinking.
b) For changing digits, press «+» to scroll up or «-» to scroll down.
c) Press «
» to confirm the selected value.
Note: If without any key press within 60 seconds or by pressing «C», the balance returns to the previous active
application.
On completion of the settings, your balance is ready for performing the density determination of liquids.
Note: Taring the balance is possible at any time.
PRESS EN
The balance prompts you: "PRESS ENTER TO START".
Press « » to start. Tare/Zero is executed.
WEIGH IN AIR
The balance prompts you to weigh the solid in air "WEIGH IN AIR".
a) Load the solid.
b) Press « » to initiate the measurement.
WEIGH IN LIQUID
The balance prompts you to weigh the solid in the auxilliary liquid
"WEIGH IN LIQUID".
a) Load the solid.
b) Press «
» to initiate the measurement.
5.988
*
The balance now shows the determined density of the solid.
Note:
This result has already been corrected for the air buoyancy. The
buoyancy caused by the two immersed wires (Ø 0.6 mm) can be
neglected.
By pressing «C», the balance returns to "PRESS ENTER TO START".
Result:
Press « », the result will be printed.
58 Application "Density"