Operating instructions
15 Application "Division Factor Weighing"
The "Division Factor Weighing" divide a predefined factor by the weight value (in grams)
(result = factor / weight) and have it rounded to a predefined number of decimal places.
Requirement: The function "FACTOR.D" must be assigned to the «F» key (see advanced
menu topic "ASSIGN:F").
FACTOR.D
Activate function "FACTOR.D" by pressing and holding the «F» key.
SET.F.DIV
001
.
0000
009
.
8
100
1
Setting the Factor Value:
a) Press «
» to execute "SET.F.DIV". Either the factor 1 appears as
default value or the factor that was saved most recently.
b) Press « » to select a digit. The selected digit is blinking.
c) For changing digits, press «+» key to scroll up or «–» to scroll
down.
d) Press « » briefly to confirm the selected factor (no automatic
acceptance).
Note: Zero for division factor value is outside the allowed range, the
error message "FACTOR OUT OF RANGE" will be displayed.
SET.STEP
0.0001
0.0050
2
Setting the step value:
"SET.STEP" appears in the display, and the program changes automat
ically to allow the display increments to be entered. The smallest pos
sible display increment appears as default value, or the last value that
was saved.
a) Press « » to execute "SET.STEP".
b) Press « » to select a digit. The selected digit is blinking.
c) For changing digits, press «+» to scroll up or «–» to scroll down.
d) Press «
» to confirm the selected step (no automatic accep
tance).
Note: The allowed range for the step depends on the factor and the
resolution of the balance. If it is outside the allowed range the error
message "STEP OUT OF RANGE" will be displayed.
Note: If without any key press within 60 seconds, the balance returns to the previous active application. Press
«C» to cancel and returns to the previous active application.
On completion of the setting procedure, your balance is ready for division factor weighing.
55
Application "Division Factor Weighing"