Operating instructions

14 Application "Multiplication Factor Weighing"
The "Multiplication Factor Weighing" application allows you to multiply the weight value (in
grams) by a predefined factor (result = factor * weight) and have it calculated to a predefined
number of decimal places.
Requirement: The function "FACTOR.M" must be assigned to the «F» key (see advanced
menu topic "ASSIGN:F").
FACTOR.M
Activate function "FACTOR.M" by pressing and holding the «F» key.
SET.F.MUL
001
.
0000
009.8100
1
Setting the factor value:
a) Press «
» to execute "SET.F.MUL". Either the factor 1 appears as
default value or the factor that was saved most recently.
b) Press « » to select a digit. The selected digit is blinking.
c) For changing digits, press «+» to scroll up or «» to scroll down.
d) Press « » to confirm the selected factor (no automatic accep
tance).
Note: Zero for multiplication factor value is outside the allowed range,
the error message "FACTOR OUT OF RANGE" will be displayed.
SET.STEP
001.0
005.0
2
Setting the step value:
"SET.STEP" appears in the display, and the program changes automat
ically to allow the display increments to be entered. The smallest pos
sible display increment appears as default value, or the last value that
was saved.
a) Press « » to execute "SET.STEP".
b) Press « » to select a digit. The selected digit is blinking.
c) For changing digits, press «+» to scroll up or «» to scroll down.
d) Press «
» to confirm the selected step (no automatic accep
tance).
Note: The allowed range for the step depends on the factor and the
resolution of the balance. If it is outside the allowed range the error
message "STEP OUT OF RANGE" will be displayed.
Note: If without any key press within 60 seconds, the balance returns to the previous active application. Press
«C» to cancel and returns to the previous active application.
On completion of the setting procedure, your balance is ready for multiplication factor weighing.
53
Application "Multiplication Factor Weighing"