Operating instructions
 13
.
45
g
M.START
TIME: 5
C
TIME: 1
 13
.
39
g
C
a)  Press « » to zero if needed.
b)  If using a container: place empty container on weighing pan and
press « » to tare the balance.
c)  Load sample weight.
d)  – If you have selected function "Manual Start" "M.START", press
«
» to start the weighing.
– If you have selected function "Auto Start" "A.START", the weigh
ing starts automatically on relative stability. For weighing samples
below 5 g the weighing must be started manually by pressing
« ».
e)  Read off result. The result of the dynamic weighing is displayed
with an asterisk (* = calculated value).
f)  Unload sample weight.
g)  "Manual Start" only, press « » to zero and go back to
"M.START". 
Note: 
●
The remaining weighing time (in seconds) is displayed continu
ously. You can cancel the countdown by pressing «C».
●
The weight value remains in the display until the sample weight is
removed from weighing pan ("Auto Start" only) or « » is
pressed.
52 Application "Dynamic Weighing"