12 Application "Totaling"
The "TOTALING" application allows you to weigh in different samples to add their weight val
ues and to totalize them. 1 to 999 samples are possible.
Requirement: The function "TOTAL" must be assigned to the «F» key (see advanced menu
topic "ASSIGN:F").
CLR.M:NO
TOTAL
CLEARED
0
.
00
g
a) Activate function "TOTAL" by pressing and holding the «F» key.
b) For a new totaling evaluation press « » (or «+» or «») to enter
"CLR.M:YES" and press « » to clear the memory.
Note: If the memory is already cleared (sample counter is 0) the
memory clear question will be not displayed.
c) Press « » to zero or tare the balance.
46
.
36
g
- 1 -
999
0.00
g
0
.
00
g
C
C
Weighing in the sample weight:
a) If using a container: place empty container on the weighing pan
and press «
» to zero or tare the balance.
b) Load the first sample weight.
c) Press « ». The display shows the sample count "- 1 -" and the
current weight is stored.
Note: When the sample counter is displayed you may press «C»
to undo (drop) this sample.
d) Unload the first sample weight. The display shows zero.
Weighing in further sample weights:
The same procedure as for the first sample weight.
1...999 samples are possible.
Results:
If the numbers of sample are greater than or equal to 2, press « »,
the results are displayed and printed.
Displayed results:
a) Press « » briefly to show the
totalized value.
b) Press «C» briefly to cancel.
0.5 seconds
number of samples
879
N
totalized value
TOTAL
8789
.
79
g
49Application "Totaling"