Operating instructions

11 Application "Formulation" (Net Total Formulation)
The "Formulation" (Net Total) application allows you to
weigh in (add and store) up to 999 individual component weights and displays the total.
If a printer is connected, the component weights are printed individually and as a total.
tare/pre-tare and store up to 999 container weights and displays the total. If a printer is
connected, the tare weights are printed out individually and as a total.
Requirement: The function "FORMULA." must be assigned to the «F» key (see advanced
menu topic "ASSIGN:F"). Connect a printer or a PC if present.
CLR.
M:NO
FORMULA
CLEARED
a) Activate function formulation "FORMULA." by pressing and holding
the «F» key.
b) Press « » to continue formulation weighing. For a new formula
tion press « » (or «+» or «-») to select "CLR.M:YES" and press
« » to clear the memory.
Note: If the memory is already cleared (sample and tare/pre-tare
counter is zero) the memory clear question will be not displayed.
#
-T1-
0
.
00
g
Net
10
.
33
g
C
0
.
00
g
#
Tare container (if used):
a) Press « » to zero or tare the balance if needed.
b) Place the empty container on the weighing pan.
c) Press « » . The container is tared and the tare count "- T1
-" is displayed and the tare weight is printed.
Note:
If you pre-tare via MT-SICS (e.g. bar code reader) "- PT1 -" is dis
played.
Zero range setting (menu topic "ZERO.RNG") has no effect. The
zero-limit is less than or equal 10d.
Net
Net
C
Weighing the first component weight:
a) Load the first component weight.
b) Press « ». The display briefly shows the component count "- 1 -
" , the current weight is stored as sample and the component
weight is printed. The display is set back to zero.
Weighing further component weights:
The same procedure as for the first component weight with the same or
new container).
1...999 sample values are possible.
max 999 tare values are possible.
max 999 pre-tare values are possible.
Results:
If the numbers of sample are greater than or equal to 2, press « »,
the results are displayed and printed.
Displayed results: 0.5 seconds
47Application "Formulation" (Net Total Formulation)