Operating instructions
11 Application "Formulation" (Net Total Formulation)
The "Formulation" (Net Total) application allows you to
●
weigh in (add and store) up to 999 individual component weights and displays the total.
If a printer is connected, the component weights are printed individually and as a total.
●
tare/pre-tare and store up to 999 container weights and displays the total. If a printer is
connected, the tare weights are printed out individually and as a total.
Requirement: The function "FORMULA." must be assigned to the «F» key (see advanced
menu topic "ASSIGN:F"). Connect a printer or a PC if present.
CLR.
M:NO
FORMULA
CLEARED
a) Activate function formulation "FORMULA." by pressing and holding
the «F» key.
b) Press « » to continue formulation weighing. For a new formula
tion press « » (or «+» or «-») to select "CLR.M:YES" and press
« » to clear the memory.
Note: If the memory is already cleared (sample and tare/pre-tare
counter is zero) the memory clear question will be not displayed.
#
-T1-
0
.
00
g
Net
10
.
33
g
C
0
.
00
g
#
Tare container (if used):
a) Press « » to zero or tare the balance if needed.
b) Place the empty container on the weighing pan.
c) Press « » . The container is tared and the tare count "- T1
-" is displayed and the tare weight is printed.
Note:
●
If you pre-tare via MT-SICS (e.g. bar code reader) "- PT1 -" is dis
played.
●
Zero range setting (menu topic "ZERO.RNG") has no effect. The
zero-limit is less than or equal 10d.
Net
Net
C
Weighing the first component weight:
a) Load the first component weight.
b) Press « ». The display briefly shows the component count "- 1 -
" , the current weight is stored as sample and the component
weight is printed. The display is set back to zero.
Weighing further component weights:
The same procedure as for the first component weight with the same or
new container).
●
1...999 sample values are possible.
●
max 999 tare values are possible.
●
max 999 pre-tare values are possible.
Results:
If the numbers of sample are greater than or equal to 2, press « »,
the results are displayed and printed.
Displayed results: 0.5 seconds
47Application "Formulation" (Net Total Formulation)