Operating instructions
10 Application "Statistics"
The "Statistics" application allows you to generate statistics of weighing values. 1 to 999
values are possible. Requirement: The function "STAT." must be assigned to the «F» key (see
advanced menu topic "ASSIGN:F"). Connect a printer or a PC if present.
CLR. M:NO
STAT
CLEARED
a) Activate function "STAT." by pressing and holding the «F» key.
b) To continue the last statistics press « ». For a new statistical
evaluation press « » (or «+», «–») to select "CLR.M:YES" and
press « » to clear the memory.
Note: If the memory is already cleared (sample counter is 0) the
memory clear question will be not displayed.
46
.
36
g
0
.
00
g
- 1 -
C
999
C
0
.
00
g
Weighing the first sample weight:
a) Press « » to zero/tare the balance if needed.
b) Load the first sample weight.
c) Press « ». The display shows the sample count "- 1 -" and the
current weight is stored as sample and the weight is printed out.
Note: When the sample counter is displayed you may press «C»
to undo (drop) this sample.
d) Unload the first sample weight.
Weighing further sample weights:
The same procedure as for the first sample weight.
●
1...999 samples are possible.
●
The next value will be accepted if the sample weight is in the range
70% –130% of the current average value. "OUT OF RANGE" will
be displayed if the sample is not accepted.
Results:
If the numbers of sample are greater than or equal to 2, press « »,
the results are displayed and printed.
Displayed results:
a) Press « » to show the next
statistical value.
b) Press «C» to cancel displaying
results and to continue weigh
ing next sample.
0.5 seconds
number of samples
5
N
average
X
50.530
g
standard deviation
S.DEV
3.961
g
relative standard deviation
S.REL
7
.
84
%
lowest value (minimum)
MIN
46
.
36
g
highest value (maximum)
MAX
55.81
g
different between the minimum
and the maximum
DIFF
9.45
g
45Application "Statistics"