Operating instructions
SET.T+
/-  %
 2
.
50
%
 20
.
28
g
2A
Step 2, setting possibility:
2A
Setting the upper and lower limits (in percentage):
a)  Press « » to start setting.
b)  Press « » to confirm the default limit of +/- 2.5 % or enter the
limit value by scrolling up («+» key) or down («–» key). Press
« » to confirm the limits.
Note: Press «
» to switch between "UNIT 1" and Unit "%". 
 SET T+/-  %
2B
 2
.
50
%
 20
.
28
g
 0
.
25
g 
 0
.
36
g
Step 2, setting possibility:
2B
Setting the upper and lower limits by weight:
a)  Press «
» to start setting.
b)  Press « » to switch to UNIT 1.
c)  Press « » to confirm the default limit or enter the limit value by
scrolling up («+» key) or down («–» key). Press « » to confirm
the limits.
 0
.
25
g
 2
.
50
%
Switching between percentage and weight unit 1 
Press « » to switch between setting in percentage and setting by
weight .
 20
.
28
g
3
 T.BEEP: Y 
Step 3:
3
Setting tolerance beep:
The tolerance beep indicates whether the weighing sample lies within
the tolerance by beeping three times.
Note: The beep level corresponds to the setting in menu topic
"STB.BEEP" (Basic menu). If "STB.BEEP" is set to "OFF", the tolerance
beep level is medium.
To activate tolerance beep press « ». To deactivate tolerance beep
press « » to select "N" and press « ».
Note:
●
If without any key press within 60 seconds, the balance returns to the previous active application. Press
«C» to cancel and returns to the previous active application.
●
The nominal weight must be at least 10 digit.
On completion of the setting procedure, your balance is ready for checkweighing. 
43Application "Check Weighing"