Operating instructions

0
.
00
g
1
PCS
2
.
74
g
1
C
------
PCS.WGT
C
Setting possibility
C
Setting the reference for one piece in weighing mode
a) Select "PCS.WGT" by scrolling with « ».
b) Press « » to tare. If needed: place empty container on the
weighing pan and tare again.
c) Add one reference piece to container. The weight of one piece is
displayed.
d) Press «
» to confirm.
Note: With approved balances, this setting is not available in selected
countries.
2
.
68
g
*
0
PCS
2
.
74
g
*
PCS.WGT
D
Setting possibility
D
Setting the reference for one piece in manual mode
a) Select "PCS.WGT" by scrolling with « ».
b) Press « » to confirm.
c) Enter the final reference one piece weight by scrolling up («+» key)
or down («-» key). Speed up by press and hold.
d) Press «
» to confirm.
Note: With approved balances, this setting is not available in selected
countries.
2
.
74
g
2
.
74
g
,
Switching between manual mode and weighing mode
Press « » to switch between manual and weighing mode.
Note: By switching from weighing mode to manual mode the weight
value will be transferred and can be changed manually.
Note: If without any key press within 60 seconds, the balance returns to the previous active application. Press
«C» to cancel and returns to the previous active application.
On completion of the setting procedure, your balance is ready for piece counting.
812
.
7
g
75
PCS
Switching between piece count and weight display.
You can use the « » key at any time to switch the display between
piece display, weighing unit "UNIT 1", "RECALL" value (if activated)
and weighing unit "UNIT 2" (if different from "UNIT 1").
Note:
The "RECALL" value is displayed with an asterisk (*) and icon "M" and can not be printed.
Take into account minimum values: min. reference weight = 10d (10 digits), min. piece weight* = 1d (1
digit)!
* with approved balances in selected countries: min 3e
The current reference weight remains stored until the reference setting is changed.
39
Application "Piece Counting"