Operating instructions
"REPEAT.T" Repeatability test (models with internal weights only)
"DISPLAY" Display test
"KEYPAD.T" Key test
"CAL.MOT.T" Motor test (models with internal weights only)
"BAL.HIST" Balance history
"CAL.HIST" Calibration history
"BAL.INFO" Balance information
"SRV.PROV" Service provider information
"SRV.ICON" – Service Reminder
This menu topic allows you to switch the service reminder "
" on or off.
"ON" Service reminder "
" switched on.You will be informed after one
Year or 8000 operating hours to call service for recalibration.
This will be indicated by the flashing service icon: " ". (Factory
setting)
"OFF" Service reminder " " switched off.
"SRV.D.RST" – Service Date Reset
This menu topic allows you to reset service date and hours.
Note: This menu topic is only available if “SRV.ICON” setting “ON” was selected.
To toggle between “YES?” and “NO?” press « » (or «+» or «-»)
6.3.4 Interface Menu
"RS232" – RS232C Interface
1)
At this menu topic you can select the peripheral device connected to the RS232C interface and specify how the
data is transmitted.
"PRINTER" Connection to a printer. (Factory setting)
"PRT.STAB" If the «
» key is pressed, the next stable weight value will be
printed. (Factory setting)
"PRT.AUTO" Every stable weight value will be printed, without pressing the
« » key.
"PRT.ALL" If the « » key is pressed, the weight value will be printed
regardless of stability.
"PC-DIR." Connection to a PC: the balance can send data (as a Keyboard)
to the PC used for PC applications e.g. Excel.
Note: The balance sends the weight value without the unit to the
PC.
"PRT.STAB" If the «
» key is pressed, the next stable weight value will be
sent followed by an enter. (Factory setting)
"PRT.AUTO" Every stable weight value will be sent followed by an enter, with
out pressing the « » key.
"PRT.ALL" If the « » key is pressed, the weight value will be sent followed
by an enter regardless of stability.
32 The Menu