Operating instructions
"RESET" – Reset Balance Settings
This menu topic allows you to cal-up the factory settings.
To toggle between "YES?" and "NO?" press « » (or «+» or «–»).
Note: A reset of the balance will not change the "DATE" and "TIME" settings.
6.3.3 Advanced Menu
"ENVIRON." – Environment Settings
This setting can be used to match your balance to the ambient conditions.
"STD." Setting for an average working environment subject to moderate
variations in the ambient conditions. (Factory setting)
"UNSTAB." Setting for a working environment where the conditions are con
tinuously changing.
"STABLE" Setting for a working environment which is practically free from
drafts and vibrations.
"CAL" – Adjustment (calibration)
In this menu topic you can preselect the function of the «
» key. Your balance can be adjusted with internal or
external weights by pressing the « » key. If you have attached a printer to your balance, the data of the adjust
ment (calibration) are printed out.
"ADJ.OFF" The adjustment is switched off. The « » key has no function.
"ADJ.INT" Internal adjustment: adjustment is performed at a keystroke with
the built-in weight (depending on the model, see technical data).
"ADJ.EXT" External adjustment: adjustment is performed at a keystroke with
a selectable external weight.
Note:This function is not available for approved balances *
(depend on selected countries' certification legislation). * except
OIML accuracy class I approved models.
"200.00 g" Defining the external adjustment weight: define the weight of
the external adjustment weight (in grams).
Factory setting: depends on the model.
"DATE.FRM" – Date Format
This menu topic allows you to preselect the date format.
The following date formats are available:
Display examples Printing examples
"DD.MM.Y" 01.02.09 01.02.2009
"MM/DD/Y" 02/01/ 02/01/2009
"Y-MM-DD" -02-01 2009-02-01
"D.MMM Y" 1.FEB.09 1.FEB 2009
"MMM D Y" FEB.1.09 FEB 1 2009
Factory setting: "DD.MM.Y"
29The Menu