Operating instructions
5 Weighing Made Simple
This section shows you how to perform simple weighings and how you can accelerate the
weighing process.
5.1 Switching the Balance On and Off
This section shows you how to perform simple weighings and how you can accelerate the weighing process.
On
Off
8888888
XXXXXXX
0
.
00
g
Switching On
a) Remove any load from weighing pan.
b) Press «On».
The balance performs a display test (all segments in the display light
up briefly), "WELCOME", Software version, Maximum load and Read
ability appears briefly. (Startup "FULL" mode only)
The balance is ready for weighing or for operation with the last active
application.
SHUTOFF
On
Off
Switching Off
Press and hold the «Off» key until "SHUTOFF" appears on the display.
Release the key.
When Quickstart is selected (Advanced menu, topic "STARTUP" > "QUICK"): Once your bal
ance has been switched off, it is in standby mode. In this case your balance needs no warm-
up time in the standby mode and is immediately ready for weighing. If you wish to perform a
weighing, you now only need to place the sample on the weighing pan and the balance
immediately displays the result. There is no need to switch it on with the «On/Off» key.
●
If your balance has been switched off after a preselected time, the display is dimly lit and
shows date, time, maximum load and readability.
●
If your balance has been switched off manually, the display is off.
Note:
●
Quickstart is not possible with approved balances (only available in selected countries).
●
Standby mode is available on line powered balances only.
22