Operating instructions

Transporting Over Short Distances
For balances with a draft shield: Observe the following instructions to transport your bal
ance over a short distance to a new location: Never lift the balance using the glass draft
shield. The draft shield is not sufficiently fastened to the balance.
Transporting Over Long Distances
If you would like to transport or send your balance over long distances, use the complete original packaging.
4.6 Weighing Below the Balance
The balances are equipped with a hanger for carrying out weighings below the work surface (weighing below
the balance).
Attention:
Do not place the balance on the pan support location bolt.
a) Switch off the balance and remove the power cable and any
interface cable from the balance.
b) Remove weighing pan, pan support and draft shield element
or "Easy draft shield" if present.
c) Turn the balance carefully on its side.
d) Remove one of the caps (A) depending on the models.
e) Then turn the balance to its normal position and simply rein
stall all components in the reverse order.
4.7 Adjustment (Calibration)
To obtain accurate weighing results, the balance must be adjusted to match the gravitational
acceleration at its location. Adjusting is necessary:
before the balance is used for the first time.
at regular intervals during weighing service.
after a change of location.
4.7.1 Adjustment with Internal Weight
Note: On models with internal weight only (see technical data).
To obtain accurate results, the balance must be connected to the power supply for
approximately,
30 minutes for balances with redability of 1 mg to 5 g
60 minutes for balances with redability of 0.01 mg to 0.1 mg
in order to reach operating temperature before adjusting.
19Setting up the Balance