10.4  Tune parameters
(PATH: MENU / PID Setup / Tune Parameters)
  This menu assigns control to a measurement and sets the setpoint, tuning parameters and non-
linear functions of the controller through a series of screens.
10.4.1  PID assignment & tuning
  Assign the measurement, a, b, c, or d to be controlled after ”PID on_”. Set the Gain (unitless), 
integral or reset time Tr (minutes) and rate or derivative time Td (minutes) needed for control. 
Press [ENTER]. Gain, reset and rate are later adjusted by trial and error based on process re-
sponse. Always begin with Td at zero.
10.4.2  Setpoint & deadband
  Enter the desired setpoint value and the deadband around the setpoint, where no proportional 
control action will take place. Be sure to include the units multiplier u or m for conductivity. 
Press [ENTER].
10.4.3  Proportional limits
  Enter the low and high proportional limits -the range over which control action is required. Be 
sure to include the units multiplier u or m for conductivity. Press [ENTER].
10.4.4  Corner points
  Enter the low and high corner points in conductivity, pH, dissolved oxygen units and the respec-
tive output values from -1 to +1, shown in the figure as -100 to +100%. Press [ENTER].