Specifications
Transmitter M400/2(X)H, M400G/2XH    64
© 03 / 2014 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland  Transmitter M400/2(X)H, M400G/2XH
Printed in Switzerland  30 031 683
  Navigate to the menu “DLI Stress Adjustment” by using the  and . keys and press [ENTER]. 
  Adjust the Process Stress parameter based on the particular application and/or requirements
Low:  DLI, TTM and ACT will be increased approximately 25% compared with “Medium”. 
Medium:  Default value, (equal DLI, TTM and ACT values based on former firmware versions of 
the transmitter). 
High:  DLI, TTM and ACT will be reduced approximately 25% compared with “Medium”.
Pressing the [ENTER] key will bring up the Save Changes dialog. Selecting No will discard the 
entered values, selecting Yes will make activate entered values.
8.7  Display
(PATH: Menu / Configure / Display) 
  Enter configuration mode as described in Section 8.1 “Enter Configuration Mode”.
This menu allows for the configuration of the values to be displayed and also the configuration 
of the display itself.
8.7.1  Measurement
The display has 4 lines. Line 1 on top and Line 4 on the bottom.
  Select the values (Measurement a, b, c or d) to be displayed on each line of the display.
The selection of the values for a, b, c, d needs to be done under Configuration / measure-
ment / Channel Setup.
  Select the ”Error Display” mode. If this is set to ”On” when an alarm or warning has occurred, 
the message ”Failure – Press ENTER” will be displayed on Line 4 when an alarm occurs in the 
normal measurement mode.
  Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will 
 discard the entered values, selecting Yes will make the entered values the current ones.