11.2.2 Calibrate analog
Select the Analog Output you wish to calibrate. Each Analog output can be calibrated at 4 and
20 mA.
Connect a Meter to the Analog output terminals and then adjust the five digit number in the
display to set the output to 4mA and repeat for 20 mA.
As the five digit number is increased the output current increases and as the number is
decreased the output current decreases. Thus coarse changes in the output current can be
made by changing the thousands or hundreds digits and fine changes can be made by
changing the tens or ones digits.
Pressing the [ENTER] key after entering both values will bring up a confirmation screen.
Selecting „No" will discard the entered values, selecting „Yes" will make the entered values
the current ones.
11.2.3 Calibrate unlock
Select this Menu to configure the CAL Menu, see chapter 7.
Selecting „Yes" means that Meter and Analog Output calibration Menus will be selectable
under the CAL Menu. Selecting „No" means that only the Sensor calibration is available under
the CAL Menu.
Pressing the [ENTER] key after the selection will bring up a confirmation screen.
11.3 Tech service
(PATH: Menu/Service/Tech Service)
C
NOTE: This Menu is for Mettler-Toledo Service personnel use only.