Operation Manual Transmitter O2 M300 Transmitter O2 M300 52 121 305
Transmitter O2 M300 © 05/07 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Printed in Switzerland Transmitter O2 M300 52 121 305
Transmitter O2 M300 Operation Manual Transmitter O2 M300 © 05/07 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Printed in Switzerland Transmitter O2 M300 52 121 305
Transmitter O2 M300 Index 1 Introduction_ ___________________________________________________________________________________________ 7 2 Safety instructions_______________________________________________________________________________________ 7 2.1 Definition of equipment and documentation symbols and designations_________________________________________ 7 2.
Transmitter O2 M300 8 Configure_____________________________________________________________________________________________ 30 8.1 Exit configuration mode_ ___________________________________________________________________________ 30 8.2 Enter configuration mode___________________________________________________________________________ 30 8.3 Measurement_ ___________________________________________________________________________________ 30 8.3.
Transmitter O2 M300 12 Info__________________________________________________________________________________________________ 54 12.1 Messages_______________________________________________________________________________________ 54 12.2 Calibration data_ _________________________________________________________________________________ 54 12.
Transmitter O2 M300 1 Introduction Statement of Intended Use – The M300 O2 transmitter is a single-channel online process instrument for measuring conductivity or resistivity of fluids. It will interface with a variety of different Mettler-Toledo sensors, which connect to the transmitter using cables of varied lengths. A large four line backlit Liquid Crystal Display conveys measuring data and setup information.
Transmitter O2 M300 The following is a list of general safety instructions and warnings. Failure to adhere to these instructions can result in damage to the equipment and/or personal injury to the operator. – The M300 Transmitter should be installed and operated only by personnel familiar with the transmitter and who are qualified for such work. – The M300 Transmitter must only be operated under the specified operating conditions (see section 16).
Transmitter O2 M300 9 3 Unit overview M300 models are available in both a 1/ DIN and 1/2DIN case size. The 1/ DIN is a panelmount only design and the 1/2DIN models provides an integral IP65 housing for wall-, or pipe-mount. 3.
Transmitter O2 M300 10 3.
Transmitter O2 M300 11 3.3 Control/Navigation Keys 3.3.1 Menu Structure Below is the structure of the M300 menu tree: 3.3.2 Navigation keys ESC Menu Cal Info Enter 3.3.2.1 Navigating the menu tree Enter the desired main Menu branch with the , or keys. Use the and keys to navigate through the selected Menu branch.
Transmitter O2 M300 12 3.3.2.2 Escape Press the and key simultaneously (escape) to return to the Measurement mode. 3.3.2.3 Enter Use the ↵ key to confirm action or selections. 3.3.2.4 Menu Press the key to access the main Menu. 3.3.2.5 Calibration mode Press the key to enter Calibration Mode. 3.3.2.6 Info mode Press the key to enter Info Mode 3.3.
Transmitter O2 M300 3.3.5 13 Navigation with ↑ in Display If a ↑ is displayed on the bottom right hand corner of the display, you can use the or the key to navigate to it. If you click [ENTER] you will navigate backwards through the menu (go back one screen). This can be a very useful option to move back up the menu tree without having to exit into the measuring mode and re-enter the menu. 3.3.
Transmitter O2 M300 4 Installation instruction 4.1 Unpacking and inspection of equipment 14 Inspect the shipping container. If it is damaged, contact the shipper immediately for instructions. Do not discard the box. If there is no apparent damage, unpack the container. Be sure all items shown on the packing list are present. If items are missing, notify Mettler-Toledo immediately 4.1.
Transmitter O2 M300 15 4.1.2 – – – – – – – – Installation procedure – 1/4DIN models Make cutout in panel (see dimensions cutout drawing). Be sure surface surrounding cutout is clean, smooth and free of burrs. Slide face gasket (supplied with transmitter) around transmitter from the back of the unit. Place transmitter into cutout hole. Be sure there are no gaps between the transmitter and panel surface.
Transmitter O2 M300 4.1.3 16 Panel cutout dimensional information – 1/2DIN models 1/2DIN Model transmitters are designed with an integral rear cover for stand-alone wall mount installation. The unit may also be wall mounted using the integral rear cover. See installation instructions in Section 4.1.4. Below are cut-out dimensions required by the 1/2DIN models when mounted within a flat panel or on a flat enclosure door. This surface must be flat and smooth.
Transmitter O2 M300 4.1.4 17 Installation procedure – 1/2DIN models For Wall Mount: – Remove rear cover from front housing. – Start by unscrewing the four screws located on the face of the transmitter, in each corner. This allows the front cover to swing away from the rear housing. – Remove the hinge-pin by squeezing the pin from each end. This allows the front housing to be removed from the rear housing – Drill out wall-mount breakthroughs in the rear housing.
Transmitter O2 M300 18 4.2 Connection of power supply All connections to the transmitter are made on the rear panel of all models. Be sure power to all wires is turned off before proceeding with the installation. High voltage may be present on the input power wires and relay wires. A two-terminal connector on the rear panel of all M300 models is provided for power connection. All M300 models are designed to operate from a 20–30 VDC or a 100 to 2 0 VAC power source.
Transmitter O2 M300 19 4.2.
Transmitter O2 M300 20 4.3 Connector PIN definition 4.3.1 TB1 and TB2 for 1/2DIN and 1/4DIN versions 4.3.2 TB3 for 1/2DIN and 1/4DIN versions Sensor Anode Colour Coax shield/red Pin no.
Transmitter O2 M300 C 21 4.4 Connection of sensor 4.4.1 Connecting the sensor to the VP cable NOTE: Be sure to observe the sensor instruction manual.
Transmitter O2 M300 4.4.2 C 22 Typical wiring NOTE: Wire colors only valid for connection with VP cable, blue and grey not connected.
Transmitter O2 M300 5 Placing transmitter in, or out, of service 5.1 Placing transmitter in service 23 After connecting the transmitter to power supply circuit, it will be active as soon as the circuit is powered. 5.2 Placing transmitter out of service First disconnect the unit from the main power source, then disconnect all remaining electrical connections. Remove the unit from the wall/panel.
Transmitter O2 M300 6 24 Quick setup (PATH: Menu/Quick Setup) While in Measurement mode press the [MENU] key to bring up the Menu selection. Select "Quick Setup" and press the [ENTER] key. The display prompts you to enter the Quick Setup security code. Enter the Quick Setup security code to "xxxxx", then press the [ENTER] key to confirm the Quick Setup security code.
Transmitter O2 M300 6.3 25 Set points After configuring the Analog Output a Set Point can be configured for that measurement. If „No" is selected and [ENTER] is pressed then the quick setup is done and the menus are exited without setting up any Setpoint. Selecting '„Yes"' and pressing [ENTER] means a Set Point can be configured for channel a.
Transmitter O2 M300 7 26 Calibration (PATH: Cal) 7.1 C Exit calibration mode NOTE: to exit Calibration mode at any time press the and key simultaneously (escape). The transmitter returns to the Measurement mode and the old calibration values remain active. 7.2 Enter calibration mode While in Measurement mode press the key. The display prompts you to enter the calibration security code.
Transmitter O2 M300 7.3.2 27 Process calibration Enter Calibration mode as described in section 7.2. Slope or Zeropt Process calibration can be selected. Choose a Slope Process calibration if you are working in the high measuring range (i.e. around 100%sat) and choose a Zero Process calibration if you are working in the low measuring range. Take a snap shot of the measuring value by pressing the [ENTER] key.
Transmitter O2 M300 7.4 28 Temperature calibration Enter Calibration mode as described in section 7.2 and select "Temperature Calibration". 7.4.1 One – point temperature calibration See 7.4 on how to enter the Temperature Calibration menu. Slope or Offset can be selected with the 1 point calibration. Select "Slope" to recalculate the Slope factor M and Offset to recalculate the offset calibration factor A. Enter Point 1 The newly calculated value – either M or A – is displayed.
Transmitter O2 M300 29 The newly calculated values M and A are displayed. Select „Yes" to save the new calibration values and the successful Calibration is confirmed on the display. 7.5 Edit Enter Calibration mode as described in section 7.2 and select "Edit". All calibration constants of the primary measurement (p) for the sensor are displayed on Line 3. All calibration constants of the secondary measurement (s) for the sensor are displayed on Line 4.
Transmitter O2 M300 8 30 Configure (PATH: Menu/Configure) 8.1 C Exit configuration mode OTE: to exit Configuration mode at any time press the and key simultaneously (escape). N The transmitter returns to the Measurement mode and the old settings remain active. 8.2 Enter configuration mode While in Measurement mode press the key. Press the or key to navigate to the Configure – Menu.
Transmitter O2 M300 8.3.1 31 Channel setup Press the [ENTER] key to select this Menu. The 4 lines of the display can now be configured with a value. Pressing the [ENTER] key will display the selection for lines c and d. Pressing the [ENTER] key again will bring up the Save Changes dialog. 8.3.2 Temperature source Press the [ENTER] key to select this Menu. The following option can be chosen: Fixed or Use this channel Fixed: allows a specific temperature value to be entered.
Transmitter O2 M300 8.3.3 32 O2 Press the [ENTER] key to select this Menu, which allows for the setting of additional measurement parameters for the O2 measurement. Atmospheric pressure and Process Pressure can be set. The default value for AtmPres is 760.0 and the default unit is mmHg. The units ProcPres and AtmPres do not have to be the same. The salinity of the measured solution and the relative humidity of the calibration gas can also be entered.
Transmitter O2 M300 8.4 33 Analog outputs (PATH: Menu/Configure/Analog Outputs) Press the [ENTER] key to select this Menu, which lets you configure the 2 Analog Outputs. The analog output number (1 or 2, default is 1), the channel (a, b, c, d or blank (none) ) and a value (off, 3.6 mA or 22.0 mA. default is off) is configured. If a mA value is set it is the value the analog output will go to if any of the alarm conditions occur The Aout type can be Normal, Bi-Linear, Auto-Range or Logarithmic.
Transmitter O2 M300 8.5 34 Set points (PATH: Menu/Configure/Set Points) Press the [ENTER] key to select this Menu. Up to 4 Set Points can be configured on one of the channels (a .. d). Possible types are Off, High, Low, Outside and Between. An 'Outside' Setpoint will cause an alarm condition whenever the measurement goes above its high limit or below its low limit.
Transmitter O2 M300 8.6 35 Alarm/Clean (PATH: Menu/Configure/Alarm/Clea) This Menu allows the configuration of Alarm and Clean functionality. 8.6.1 Alarm Select the Relay to be used for the Alarm. Select either the "Power failure" or "Software failure" event to be alarmed. If either of these is set to yes then the alarm will come on and an alarm message will be recorded if: 1. there is a power failure or power cycling 2.
Transmitter O2 M300 8.6.2 36 Clean Configure the Relay to be used for the cleaning cycle. The Cleaning Interval can be set from 0.000 to 999.9 hours. Setting it to 0 turns the clean cycle is off. The cleaning time can be 0 to 9999 seconds and must be smaller than the Cleaning Interval. Select the desired Relay state: Normal or Inverted. Pressing the [ENTER] key again will bring up the Save Changes dialog.
Transmitter O2 M300 8.7.2 37 Resolution This menu allows the setting of the resolution of each displayed value. Possible settings are 1, 0.1, 0.01, 0.001 or Auto. Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will discard the entered values, selecting „Yes" will make the entered values the current ones. 8.7.3 Backlight This Menu allows the setting of the back light options of the display. Possible settings are On, On 50% or Auto Off 50%.
Transmitter O2 M300 8.8 38 Hold outputs (PATH: Menu/Configure/Hold Outputs) This Menu allows for the configuration of the Hold status. The HOLD function is used if a Easy Clean system is controlled from the M300 or if a control of the HOLD functionality is required from an overlying system. Initiating a hold condition will maintain the analog signal output and relay status at the value/ state at the time the Hold is initiated, for as long as the Hold state is maintained.
Transmitter O2 M300 9 39 System (PATH: Menu/System) While in Measurement mode press the key. Press the or key to navigate to the System – Menu. After selecting the System – Menu enter the System security code and then press the [ENTER] key to confirm the code. 9.1 Set language (PATH: Menu/System/Set Language) This Menu allows the configuration of the Display language. Following Selections are possible: English, French, German, Italian and Spanish.
Transmitter O2 M300 40 9.3 Passwords (PATH: Menu/System/Passwords) This Menu allows for the configuration of Operator and Administrator Passwords, as well as setting up a List of allowed Menus for the Operator. The Administrator has rights to access all Menus. The Passwords Menu is protected: Enter the Administrator Password to enter the Menu. 9.3.1 Changing passwords See 9.3 on how to enter the Passwords Menu.
Transmitter O2 M300 41 9.4 Set/clear lockout (PATH: Menu/System/Set/Clear Lockout) The user will be asked for a password before being allowed into any menus if the Lockout functionality is enabled. The Lockout – Menu is protected: Enter the Administrator Password and select „Yes" to enable or „No" to disable the Lockout functionality. Pressing the [ENTER] key after the selection will bring up the Save Changes dialog. 9.
Transmitter O2 M300 42 9.5.2 Reset meter calibration This Menu allows the reset of the meter's calibration factors to the last factory calibration values. Pressing the [ENTER] key after the selection will bring up a confirmation screen. Selecting „No" will discard the entered value, selecting „Yes" will make the entered value the current one. 9.5.3 Reset analog calibration This Menu allows the reset of the Analog Output calibration factors to the last factory calibration values.
Transmitter O2 M300 43 10 PID setup (PATH: Menu/PID Setup) PID control is proportional, integral and derivative control action that can provide smooth regulation of a process. Before configuring the transmitter, the following process characteristics must be identified.
Transmitter O2 M300 44 Controller with Corner Points Proportional limit +100% value 120% 100% 80% 60% Direct corner point (value, %) 40% Set point value 20% Deadband + value 0% Deadband – value –20% Reverse corner point (value, %) –40% –60% –80% –100% Proportional limit –100% value –120% 0 2 4 6 8 10 12 14 Process Variable 10.1 Enter PID Setup While in Measurement mode press the key. Press the or key to navigate to the PID Setup – Menu.
Transmitter O2 M300 45 10.3 Mode (PATH: MENU/PID Setup/Mode This menu contains the selection of control modes using relays or analog outputs. Press [ENTER]. This menu assigns a relay or analog output for PID control action as well as details of their operation. Based on the control device being used, select one of the following three paragraphs for use with solenoid valve, pulse input metering pump or analog control. Pulse Length – If using a solenoid valve, select "Relays" and "PL", Pulse Length.
Transmitter O2 M300 46 10.3.1 Tune Parameters (PATH: MENU/PID Setup/Tune Parameters This menu assigns control to a measurement and sets the setpoint, tuning parameters and nonlinear functions of the controller through a series of screens. 10.3.2 PID Assignment & Tuning Assign the measurement, a, b, c, or d to be controlled after "PID on_". Set the Gain (unitless), Integral or Reset time Tr (minutes) and Rate or Derivative time Td (minutes) needed for control. Press [ENTER].
Transmitter O2 M300 47 10.4 PID Display Setup (PATH: MENU/PID Setup/PID Display Setup This screen enables display of PID control status including % output and Auto/Manual in the normal measurement mode. (In addition, for the display to be enabled, a measurement must be assigned under Tune Parameters and a relay or analog output must be assigned under Mode.) If „Yes" is selected the status (Man or Auto) and control output will be displayed on the bottom line.
Transmitter O2 M300 48 11 Service (PATH: Menu/Service) While in Measurement mode press the key. Press the or key to navigate to the Service – Menu. After selecting the Service – Menu enter the Service security code "xxxxx" and then press the [ENTER] key to confirm the code. 11.
Transmitter O2 M300 49 11.1.1 Model/Software Revision Essential information for every Service call is the model and software revision number. This Menu shows the transmitter part number, serial number and software version number. Press [ENTER] to exit from this display. 11.1.2 Digital Input The digital Input menu shows the state of the digital input. 11.1.3 Display All pixels of the display will be lit for 15 seconds to allow troubleshooting of the display.
Transmitter O2 M300 50 11.1.5 Memory If Memory is selected then the transmitter will perform a RAM and ROM memory test. Test patterns will be written to and read from all RAM memory locations. The ROM checksum will be recalculated and compared to the value stored in the ROM. 11.1.6 Set relay The Set Relays diagnostic menu allows for the manual activation/deactivation of each Relay. 11.1.7 Read relays The Read Relays diagnostic menu shows the state of each Relay.
Transmitter O2 M300 51 11.1.8 Set analog outputs This menu enables the user to set all analog outputs to any mA value within the 0-22 mA range. 11.1.9 Read analog outputs This menu shows the mA value of the analog Outputs. 11.2 Calibrate (PATH: Menu/Service/Calibrate) This menu has the options to calibrate the transmitter and the analog outputs and also allows the unlocking of calibration functionality. 11.2.1 Calibrate meter Channel A calibration can be selected as "Current" or "Temperature".
Transmitter O2 M300 11.2.1.1 52 Current Enter the value for Point 1 in milliamps of the current source connected to the input. Enter the value for Point 2 in milliamps of the current source connected to the input. Pressing the [ENTER] key after entering Point 2 will bring up a confirmation screen. Selecting „No" will discard the entered values, selecting „Yes" will make the entered values the current ones. 11.2.1.
Transmitter O2 M300 53 11.2.2 Calibrate analog Select the Analog Output you wish to calibrate. Each Analog output can be calibrated at 4 and 20 mA. Connect a Meter to the Analog output terminals and then adjust the five digit number in the display to set the output to 4mA and repeat for 20 mA. As the five digit number is increased the output current increases and as the number is decreased the output current decreases.
Transmitter O2 M300 12 54 Info (PATH: Info) Pressing the key will display the Info Menu with the options Messages and Calibration Data 12.1 Messages (PATH: Info/Messages) The most recent message is displayed. The up and down arrow keys allow scrolling through the last four messages that have occurred. Clear Messages clears all the messages. Messages are added to the message list when the condition that generates the message first occurs.
Transmitter O2 M300 12.3 55 Model/software revision Selecting Model/Software Revision will display the installed Firmware revision. The displayed information is important for any Service call.
Transmitter O2 M300 13 56 Maintenance The device contains no user repairable components. To remove dust, dirt and spots, the external surfaces of the transmitter may be wiped with a damp, lint-free cloth. 13.1 For technical support For technical support and product information contact your local METTLER TOLEDO Sales Office or representative 13.2 Front panel cleaning Clean the front panel with a damp soft cloth (water only, no solvents). Gently wipe the surface and dry with a soft cloth.
Transmitter O2 M300 14.1 57 Warning- and alarmlist Warnings Warning O2 Slope > 90nA Warning O2 Slope < 35nA Warning O2 ZeroPt > 0.3nA Warning O2 ZeroPt < –0.3nA Description Slope too big Slope too small Zero offset too big Zero offset too small Alarms Watchdog time-out Error O2 Slope > 110 nA Error O2 Slope < 30 nA Error O2 ZeroPt > 0.6 nA Error O2 ZeroPt < –0.6nA Description SW/System fault Slope too big Slope too small Zero offset too big Zero offset too small 14.
Transmitter O2 M300 16 Specifications 16.1 General specifications 58 Watchdog time-out Error O2 Slope > 110 nA Error O2 Slope < 30 nA Error O2 ZeroPt > 0.6 nA Error O2 ZeroPt < –0.
Transmitter O2 M300 16.3 Mechanical specifications for 1/4DIN version Dimensions (housing – H x W x D)* Front bezel – (H x W) Max. depth Weight Material Ingress rating * H=Height, W=Width, D=Depth 16.4 96 x 96 x 140 mm (1/4DIN model) 102 x 102 mm 125 mm (excludes plug-in connectors) 0.6 kg (1.5 lb) ABS/polycarbonate IP 65 (front)/IP 20 (housing) Mechanical specifications for 1/2DIN version Dimensions (housing – L x H x W)* Front bezel – H x W Max.
Transmitter O2 M300 17 60 Default tables Parameter Alarm (except flow) Clean Language Passwords All Relays (unless otherwise specified) Lockout Channel A Ccal constants Analog Out All analog out O2 Temperature Set point 1 Relay 3 Set point 2 Parameter Relay 4 V polarization Resolution © 05/07 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Printed in Switzerland Sub parameter Relay Diagnostics SW HW Delay Hysteresis State Relay Interval Clean time State Delay Hysteresis Administrator Ope
Transmitter O2 M300 18 61 Waranty METTLER TOLEDO warrants this product to be free from significant deviations in material and workmanship for a period of one year from the date of purchase. If repair is necessary and not the result of abuse or misuse within the warranty period, please return by freight pre-paid and amendment will be made without any charge. METTLER TOLEDO's Customer Service Dept. will determine if the product problem is due to deviations or customer abuse.
Transmitter O2 M300 62 Notes: © 05/07 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Printed in Switzerland Transmitter O2 M300 52 121 305
Transmitter O2 M300 63 Notes: © 05/07 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Printed in Switzerland Transmitter O2 M300 52 121 305
Transmitter O2 M300 64 Addresses © 05/07 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Printed in Switzerland Transmitter O2 M300 52 121 305