Specifications
Transmitter M300 FLOW 40
© 09 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter M300 FLOW
Printed in Switzerland 52 121 319
8.4 Set Points
(PATH: Menu/Configuration/Set Points)
Press the [ENTER] key to select this Menu. This menu is used to configure Setpoints.
Up to 4 Set Points for single channel transmitters and 8 for four channel transmitters can be
configured on one measurement on this screen. Types are Off, High, Low, Outside, Between and
Total Flow.
Off (Set Point is Off)
High (High value must be set)
Low (Low value must be set)
Outside (High and Low value must be set)
Between (High and Low value must be set)
Total Flow (Available only if units of total flow are chosen. A Total Flow value must be set)
Enter the desired value(s) for the Set Point and press [ENTER].
This screen provides the option to configure a setpoint to be active on an over range condition.
Select the setpoint (1 thru 4) and ”Yes” or ”No”. Select the desired relay that will activate when
the setpoint alarm condition is reached.
Over Range
Configure whether an over range should also be alarmed and which relay should be used. Once
configured, the selected relay will be activated if a sensor over-range condition is detected on the
assigned input channel.
Delay
Enter the delay time in seconds. A time delay requires the setpoint to be exceeded continuously
for the specified length of time before activating the relay. If the condition disappears before the
delay period is over, the relay will not be activated.
Hysteresis
Enter the hysteresis as a percentage-value. A hysteresis value requires the measurement to
return within the setpoint value by a specified percentage before the relay is deactivated.
For a high setpoint, the measurement must decrease more than the indicated percentage
below the setpoint value before the relay is deactivated. With a low setpoint, the measurement
must rise at least this percentage above the setpoint value before the relay is deactivated. For
example, with a high setpoint of 100, when this value is exceeded, the measurement must fall
below 90 before the relay is deactivated.