Specifications

Transmitter M300 FLOW 34
© 09 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter M300 FLOW
Printed in Switzerland 52 121 319
7.3 Edit
The Edit function is the most commonly used calibration method for flow sensors.
Enter Calibration mode as described in section 7.1 and select the channel (four channel models
only) and select Edit.
Press [ENTER] to display all calibration constants for the sensor. The calibration constants
can be changed in this menu. If the sensor type previously selected was either High or Low, the
M and A values will be displayed. If Type 2 was selected a table of K and F values will be
displayed.
Press [ENTER] until prompted to save calibration values
Select Yes to save the new calibration values and the successful Calibration is confirmed on the
display.
7.4 Verify
Enter Calibration mode as described in section 7.1 and select the channel (four channel models
only) and Verify.
The measurement value and the frequency (Hz) are shown. The meter calibration factors are
used when calculating the measurement value.
Press [ENTER] to return to the Measurement mode.