7.2  Sensor Calibration 
 This feature provides the ability to perform a One-point or Two-point Sensor flow calibration and 
”Edit” or ”Verify” saved calibration constants. The most common method of calibration for flow 
sensors is to enter the calibration constants appropriate for the sensor using the Edit function. 
Some users may choose to perform an in-line calibration using a One-point or Two-point Sensor 
flow calibration. This requires an external reference system. When performing an in-line cali- 
bration on a flow sensor, results will vary depending on the methods and calibration apparatus 
used to perform the calibration. 
Select the channel (four channel models only) and the desired calibration option. Choices are 
GPM, meters/hour or liters/minute (for a one-pt or two-pt flow calibration), Edit and Verify. Press 
[ENTER] 
7.2.1  One point calibration 
Select 1 point Calibration by pressing the  or  key followed by the [ENTER] key.
Enter the Value of calibration Point 1 from the external reference system and press the [ENTER] 
key to start calibration. The value in the 2
nd
 text line is the actual measured value from the 
sensor prior to calibration. 
After the calibration, the slope (Multiplier) calibration factor M and the offset (Adder) calibration 
factor A are displayed. 
Select Yes to save the calibration values and the successful Calibration is confirmed on the 
display. Select No to discard the entered calibration. The meter retains the old calibration values 
and returns to the Measurement mode.