Specifications
Transmitter M300 FLOW    29
© 09 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland  Transmitter M300 FLOW
Printed in Switzerland  52 121 319
6.4  Measurement Selection
Select the desired display line (a or c) of the single channel transmitter to configure the values 
to be displayed and whether this value will have an Analog Output. 
Convention (single channel models): 
1
st
 line on display  => a
3
rd
 line on display  => c
Example:
By selecting a and GPM as units, the flow rate value will be displayed on the 1
st
 line. 
By selecting c and Gals as units, the total flow value will be displayed on the 3
rd
 line of the 
display which has high resolution. 
Selecting None means the display will be blank for the line selected.
Four channel transmitters, configure the channel followed by the measurement units. 
Convention (four channel models only):
1
st
 line on display  => Channel A
2
nd
 line on display => Channel B
3
rd
 line on display => Channel C
4
th
 line on display  => Channel D
6.5  Analog Outputs
By selecting Analog Output Yes on the previous screen a linear 4–20 mA analog output Aout will 
be setup for the measurement when [ENTER] is pressed. Selecting No means that no analog 
output is setup. Aout min and Aout max are the minimum and maximum measurement values 
for the 4 and 20mA values respectively.
Analog Output defaults for Quick Setup:
Measurement a  => Aout 1
Measurement c => Aout 2
Measurement A  => Aout 1
Measurement B  => Aout 2
Measurement C  => Aout 3
Measurement D  => Aout 4