OIML Load Cell Specifications
Rated Capacity of Load
Cell (kg)
220, 550, 1,100, 2,200, 4,400 9,072, 13,608 20,412
Rated Output
1.94 ± 0.002 mV/V 2.0 ± 0.005 mV/V 2.0 ± 0.005 mV/V
Zero Balance
(% of rated output)
1.0 1.0 1.0
Combined Error Due to
Non-Linearity and
Hysteresis
0.017% of R.C. 0.02% of R.C. 0.02% of R.C.
Non-Repeatability 0.01% of R.C. 0.01% of R.C. 0.01% of R.C.
Temperature
Compensation
-10° to +40° C -10° to +40° C -10° to +40° C
Terminal Resistance
Input: 385
Signal: 350 ±2
Input: 380 ±20
Signal: 350 ±2
Input: 2,200 ±100
Signal: 2,200 ±20
Excitation Voltage 15 VDC maximum 15 VDC maximum 15 VDC maximum
Insulation Resistance
5 Giga minimum @ 50 VDC 5 Giga minimum @ 50 VDC 5 Giga minimum @ 50 VDC
Maximum Loads
% of R.C.
Safe Overload: 150
Ultimate Overload: 300
Safe Side Load: 100
Safe Overload: 150
Ultimate Overload: 300
Safe Side Load: 100
Safe Overload: 150
Ultimate Overload: 300
Safe Side Load: 100