Operating instructions
6
3.2 Preparing the balance for mounting the LP16
Removing the weighing pan
• Take off weighing pan (1) and pan support (2).
Installing the compensation weight on the LP16
If the LP16 is mounted on a PM600, PM2000, PM4000 or PM4800, the compensation weight in
the draft shield must be changed.
• Unscrew 3 screws (5); disassemble draft shield.
• Take out compensation weight (7) built in as standard.
• Install compensation weight.
• Reassemble draft shield. Marking (6) and cutout (8) must be aligned.
Resetting the operating voltage
Warning
• First ensure that the instrument is disconnected from the power supply. If the power cable (1)
is attached, the interior of the balance is live even if the display is blank!
•When the operating voltage is reset, the microfuse must be changed.
(Section 7.1)
Resetting
• Remove weighing pan, unscrew screw (2) and carefully lift off top housing (3). Reset voltage
selector (4) with a screwdriver.
• Carefully remount top housing and insert and tighten screw (2).
Selecting the location
The correct location should be selected according to the following criteria:
• Stable, vibration-free position as horizontal as possible
• No drafts