Operating instructions

25
7. Maintenance
7.1 Changing the fuses
Changing “3.15 A slow-blow” fuse
Disconnect power cable of the dryer!
Lever out holder (1) of the microfuse using a screwdriver.
Change 3.15 A slow-blow microfuse (spare in holder).
Plug in connector cable.
Changing “200 mA slow-blow” or “100 mA slow-blow” fuse
Disconnect power cable of the dryer!
Take out holder (2) particular microfuse with a screwdriver turning it to the left.
Replace “200 mA slow-blow” or “100 mA slow-blow” microfuse (spare in the accessories).
Push in holder (2) using a screwdriver and then secure by turning to right (up to stop).
Plug in power cable.
Replacing microfuses of the balance
The spare fuse is in the fuse holder (4).
Fuse ratings: 115 V = 125 mA slow-blow
230 V = 63 mA slow-blow
Disconnect power cable (3).
Take out fuse holder (4) using a screwdriver.
Remove blown fuse and insert new one.
Replace fuse holder. Plug in power cable.
2
1
3
4