Operating instructions

23
Possible cause
Connection dryer - balance interrupted
•A drying operation has been performed
beforehand
Drying operation has not been started
The calculation mode ATRO value has
been selected for the drying operation
Control bar of the balance or the STOP
key on the dryer has been pressed
•Weight value has become negative
The set abort criterion is too large, drying
is aborted since the effective weight loss
of the sample per time interval is less than
the abort criterion
Balance not yet stable
Sample evaporates even before the start
of drying
Stability not reached up to timeout since
unstable ambient conditions
Rectification
Set up correct connection
Open drying cover
Start drying
Percentage display does not appear until
after end of drying operation
Perform drying again
Perform drying again
Enter lower value and repeat drying
On stability, the drying operation starts
automatically
Set vibration adapter to unstable surround-
ings (LP16)
Increase integration time (LJ16)
Error/Display
Drying operation does not start and no error
message in the display
No switching between weight and percent-
age display possible
Termination of the drying operation
When the START key is pressed, the dashed
wait line - - - - - - appears.