Operating instructions
23
Possible cause
• Connection dryer - balance interrupted
•A drying operation has been performed
beforehand
• Drying operation has not been started
• The calculation mode ATRO value has
been selected for the drying operation
• Control bar of the balance or the STOP
key on the dryer has been pressed
•Weight value has become negative
• The set abort criterion is too large, drying
is aborted since the effective weight loss
of the sample per time interval is less than
the abort criterion
• Balance not yet stable
• Sample evaporates even before the start
of drying
• Stability not reached up to timeout since
unstable ambient conditions
Rectification
• Set up correct connection
• Open drying cover
• Start drying
• Percentage display does not appear until
after end of drying operation
• Perform drying again
• Perform drying again
• Enter lower value and repeat drying
• On stability, the drying operation starts
automatically
• Set vibration adapter to unstable surround-
ings (LP16)
• Increase integration time (LJ16)
Error/Display
Drying operation does not start and no error
message in the display
No switching between weight and percent-
age display possible
Termination of the drying operation
When the START key is pressed, the dashed
wait line - - - - - - appears.