Operating instructions

22
Error/Display
Lamp of the MODE key flashes (LP16 only)
Drying unit does not heat
The parameters to be changed can not be
called up
No entry possible
When the START key is pressed, “Err 0”
lights up (no start possible)
When the START key is pressed, “Err 1”
lights up (no start possible)
When the START key is pressed, “Err 2”
lights up (no start possible)
“Err 3”
Rectification
Close entry with ENTER
Plug in connector properly
Inform METTLER TOLEDO customer
service
Close entry with ENTER
Tare or open drying cover
Close entry with ENTER
Remove sample, tare and add sample
again
Select larger sample
Close cover
Perform new drying
Perform new drying
Possible cause
Entry not closed with ENTER
Flat connector from top part to balance not
plugged in
Infrared radiators faulty
Previous entry not closed with ENTER
•A drying operation has been performed
beforehand
Connection cable balance - dryer missing
Previous entry not closed with ENTER
Sample has been tared
•Weight of sample too low (less than 100
digits)
Drying cover opened
Percent value larger than display capacity
In the calculation of an ATRO value, the
weight value is 0