5.2.1 Automatic switch-off criterion
The LP16 and LJ16 have an automatic switch-off criterion.
The adjacent curve shows the typical weight profile of a sample during a drying operation.
The automatic switch-off criterion offers the user the possibility of terminating drying
automatically when the weight loss per time interval is less than a preset value.
Example: Weight loss < 2 mg/30 sec.
LP16: Variable, freely selectable, automatic switch-off criterion
(1 ... 99 mg/10 sec, 1 ... 99 mg/30 sec, 1 ... 99 mg/120 sec).
LJ16: fixed, automatic switch-off criterion (2 mg/30 sec).
Dry
Wet
Sample
weight
Time