Operating instructions

15
5.2 Operation of the LP16
Controls
(1) Control bar of the balance (tare bar)
(2) Keypad of the dryer
Switching display on/off
Press control bar briefly; all display segments light up for a few seconds.
The display then shows 0.000 g.
Raise control bar briefly; the display blanks out.
Taring
•Mount aluminum sample pan on the weighing pan.
Press control bar briefly; the display changes to 0.000 g.
Calibrating
Before work is performed for the first time with the LP16 dryer, the balance should be calibrated.
To achieve a correct calibration, it is advisable to attach the instrument to the power supply
30 minutes before calibrating. Calibration is necessary only if the absolute sample weight is of
importance.
The calibration procedure is described in the operating instructions of the PM balance used.