Operating instructions

14
Changing weighing speed or reproducibility
The LJ16 can be adapted to the ambient conditions. For this, the appropriate setting should be
selected for the integration time and the stability detector.
Stability detector
Press control bar until -ASd- appears, then release.
Press control bar briefly; the display changes to the next setting.
At the desired setting, wait until the display returns to the operating mode.
Integration time
Possible settings
Int-1- With very stable, vibration-free surroundings
Int-2- Normal setting
Int-3- With unfavorable ambient conditions
Performing the settings
Press control bar until -Int- appears in the display, then release.
Press control bar briefly; the display changes to the next setting.
At the desired setting, wait until the balance returns to the operating mode.
Optimum settings, see section 3.3.
Weighing speed
very fast
relatively slow
Reproducibility
good
very good
ASd -1-
ASd -2-
ASd -3-
ASd -4-
Setting stability detector