Operating instructions
30
The MT-SICS commands listed below is a selected list of available commands. For additio-
nal commands and further information please refer to the Reference Manual “MT-SICS Basic-
S balances 11780447” downloadable from the Internet under www.mt.com/sics-classic.
S – Send stable weight value
Command  S  Send the current stable net weight value.
SI – Send value immediately
Command  SI  Send the current net weight value, irrespective of 
balance stability.
SIR – Send weight value immediately and repeat
Command  SIR  Send the net weight values repeatedly,   
      irrespective of balance stability.
Z – Zero
Command  Z  Zero the balance.
@ – Reset
Command  @  Resets the balance to the condition found after 
      switching on, but without a zero setting being   
      performed.
SR – Send weight value on weight change (Send and Repeat)
Command  SR  Send the current stable weight value and then   
      send continuously the stable weight value after  
      every weight change.
      The weight change must be at least 12.5 % of  
      the last stable weight value, minimum = 30d.
ST – Send stable weight after pressing ± (transfer) key
Command  ST  Inquiry of actual status of the ST function.
SU – Send stable weight value with currently displayed unit
Command  SU  As the “S” command, but with the currently   
      displayed unit.