Operating instructions
30
The MT-SICS commands listed below is a selected list of available commands. For additio-
nal commands and further information please refer to the Reference Manual “MT-SICS Basic-
S balances 11780447” downloadable from the Internet under www.mt.com/sics-classic.
S – Send stable weight value
Command S Send the current stable net weight value.
SI – Send value immediately
Command SI Send the current net weight value, irrespective of
balance stability.
SIR – Send weight value immediately and repeat
Command SIR Send the net weight values repeatedly,
irrespective of balance stability.
Z – Zero
Command Z Zero the balance.
@ – Reset
Command @ Resets the balance to the condition found after
switching on, but without a zero setting being
performed.
SR – Send weight value on weight change (Send and Repeat)
Command SR Send the current stable weight value and then
send continuously the stable weight value after
every weight change.
The weight change must be at least 12.5 % of
the last stable weight value, minimum = 30d.
ST – Send stable weight after pressing ± (transfer) key
Command ST Inquiry of actual status of the ST function.
SU – Send stable weight value with currently displayed unit
Command SU As the “S” command, but with the currently
displayed unit.