Operating instructions
22
4.3.12  Send format (12th menu option)
Note: This menu option is only available if you have selected the setting “Host” in the 11th menu option 
(Peripheral unit)!
At this menu option you specify the data transfer format.
“SICS”:  The  MT-SICS  data  transfer  formats  are  used.  Please  refer  to  the  “MT-SICS  Reference  Manual  
Basic-S balances 11780447” downloadable from the Internet (www.mt.com/sics-classic). More 
Information please find in the Section 6.3.
“PM”*:  The following PM balance data transfer formats are used:
  S. Stb: /////1.67890/g  S. Cont:  S////1.67890/g
    SD///1.39110/g
* unidirectional, no MT-SICS commands are accepted.
4.3.13  Send mode (13th menu option)
Note: This menu option is only available if you have selected the setting “Host” in the 11th menu option 
(Peripheral unit)!
At this menu option you specify how data is transmitted to a peripheral device.
S. oFF  Send mode switched off
S. Stb  The next possible stable value will be transferred after the «±» key has been triggered.
S. Cont  All values are transferred automatically.
S. oFF
“
S. Stb
S. Cont
“
SICS
“
PŸÖ