Operating instructions

19
4.3.5 Functions (5th menu option / Operating Section 5)
In addition to simple weighing, the following functions can be selected:
F nonE No function, simple weighing (factory setting)
F count Piece counting
F 100 % Percent weighing
4.3.6 Weighing mode (6th menu option)
This setting allows you to adapt the balance to the weighing mode. Select “Std” (standard) for all normal
weighing processes or “doS” (dispensing) for dispensing liquid or powdery weighing samples. In this setting,
the balance responds very quickly to smallest weight changes.
When set to “robuSt” (absolute weighing), the balance responds only to greater weight changes, and the
weighing result is very stable.
4.3.7 Vibration adapter (7th menu option) (only available on JB-C/FACT and JB-G/FACT models)
The vibration adapter can be used to match your balance to the ambient conditions (vibrations, drafts at
location). If you work in surroundings which are practically free from temperature fluctuations, drafts, and
vibrations, select “StabLE”. On the other hand, if you work in surroundings where the conditions are constantly
changing, select “unStAbLE”. For normal ambient conditions, select “UnivErS” (Standard). This is the factory
setting.