4.3.3 FACT (3rd menu option) (only available on JB-C/FACT and JB-G/FACT models)
In this menu option you can switch the fully automatic internal adjustment (calibration) FACT on or off.
FACT switched on
This is the factory setting. The balance adjusts (calibrates) itself fully automatically.
FACT switched off
Note: With the certified versions of balances, this option can not be selected, i.e. FACT is always active
4.3.4 Protocol (4th menu option) (only available on JB-C/FACT and JB-G/FACT models)
This menu item can be used to switch the printing of the FACT fully automatic adjustment (calibration) on or
off.
Note: This menu item can only be selected if FACT is switched on and does not affect the printing of adjust-
ments with an internal or external adjustment weight (Section 4.3.3).
Protocol switched on
A record is printed out after every automatic adjustment of the balance (FACT).
Note: The protocol is printed out without a line for signatures.
Protocol switched off
This is the factory setting. If the balance adjusts automatically (FACT), a protocol is not printed out.
FACt on
FACt off
AutoCal
Prot on
Prot off
AutoCal