Operating instructions
12
Adusting with external weight
➞ To carry out this operation, in the second menu option (Adjustment) select “CAL E” (=factory setting)
(Section 4.1).
➞ Have required adjustment weight ready.
➞ Unload weighing pan
➞ Press and hold the «Cal/Menu» key down until “CAL” appears in the display, then release key. The re-
quired adjustment weight value flashes in the display.
➞ Place adjustment weight in center of pan. The balance adjusts itself automatically.
➞ When “0.00 g” flashes, remove adjustment weight.
The adjusting is finished when the message “CAL done” appears briefly in the display, followed by
“0.00 g”. The balance is again in weighing mode and ready for operation.
Notes
• JB-C/FACT and JB-G/FACT models have an internal adjustment weight and adjust themselves automati-
cally 2 times withhin 2 hours of connection to the power supply, thereafter periodically.
• Because of certification legislation, the certified JB-L-G models cannot be adjusted by the user.
• The adjustment procedure can be terminated at any time with the «C» key. The message “Abort” appears
briefly to confirm that adjustment has been canceled, and the balance reverts to weighing mode.
CAL
CAL donE
2000.00
------
0.00
------
0.00 g