10a. Stability Detect Sub-
block
The stability detection feature determines when a no-motion condition exists on the
scale. The sensitivity level determines what is considered stable. Printing and tare
operations will wait for scale stability before carrying out a command.
Stability detection occurs over a predefined period of time and allows a predetermined
acceptable amount of motion in scale divisions. The acceptable amount of motion is
considered the range and the period of time is called the interval. When the JAGXTREME
terminal is used with an IDNET base, there are four settings of stability filtering available.
To configure the sub-block:
1. Press ENTER at the Stability Detect prompt and again at the Stability Range?
prompt.
2. At the Range? prompt, enter the acceptable motion range (+/- 0.1 to 9.9
divisions).
3. Press ENTER at the Stable Readings? prompt to configure the stability parameters.
4. At the Intrval? prompt, enter the number of seconds (0 to 9.9) that the weight must
remain within the range values for a no-motion condition.
5. Continue to the next sub-block or exit the setup mode.
10b. Stability Detect Sub-
block for MMR (IDNET)
Bases
Stability Detection of an MMR (IDNET) base is determined by the base electronics. The
stability detection feature determines when a no-motion condition exists on the weighing
platform. The sensitivity level determines what is considered stable. Printing and tare
operations will wait for scale stability before carrying out the command.
The only prompt that will appear will be STABILITY ? X. Use the SELECT key to change
the value of the stability filter from 0 (no stability detection) or 1 (fast indication of
weight) through 4 (slow indication of weight).
11. Beeper Operation Sub-
block
The JAGXTREME terminal is capable of making an audible beep each time a key is
pressed and an error alarm when an inappropriate key is pressed. This sub-block lets
you enable or disable the keyboard sound.
To configure the sub-block:
1. Press ENTER at the Beeper Operation prompt to open the sub-block.
2. At the Key Beeper? prompt, select Y(es) or N(o) to enable or disable sound
when a key is pressed, as well as the double beep acknowledgment message.
3. At the Alarm Beeper prompt, select Y(es) or N(o) to enable or disable an
alarm that sounds when an error occurs or an inappropriate button is pressed.
4. Continue to the next sub-block or exit the setup mode.
Stability
Detection
Stable readings
entry
Stability range
entry
Beeper
Operation
Alarm beeper
Y(es) or No
Key beeper
Y(es) or No
Stability Detect
STABILITY ? X