Technical data

METTLER TOLEDO JAGXTREME Terminal Technical Manual
(9/00) 3-24
To configure the Tare Operation sub-block:
1. Press ENTER at the Tare Operations prompt to open the sub-block.
2. At the Enable Tare? prompt, select Y(es) or N(o) to enable or disable tare. If you
select N(o), the terminal proceeds to the Zero Operation sub-block. Access to other
tare features is not possible if tare is disabled.
3. At the Tare Interlck? prompt, select Y(es) or N(o) to enable or disable tare
interlock.
4. At the Enabl PB Tare? prompt, select Y(es) or N(o) to enable or disable
pushbutton tare.
5. At the Enabl KB Tare? prompt, select Y(es) or N(o) to enable or disable keyboard
tare.
6. At the Enb Auto Tare? prompt, select Y(es) or N(o) to enable or disable auto tare.
If Auto Tare Is Enabled
1. Press ENTER at the Tare Threshold prompt; at the Wgt? prompt, enter the desired
threshold value. The threshold value is a unit value such as 1.5 pounds. When
weight on the platform exceeds the threshold value, then settles to no-motion, the
terminal automatically tares.
2. Press ENTER at the Reset Threshold prompt. At the Wgt? prompt, enter the desired
reset threshold value. This is a unit value and must be less than the tare threshold.
When weight on the platform falls below the reset threshold value, as when the
load has been removed, the terminal automatically rearms the auto tare trigger.
3. At the Check Motion? prompt, select Y(es) or N(o) to enable or disable the motion
check. If enabled, the terminal checks for stability of the load on the platform before
resetting the auto tare trigger.
4. At the Auto Clr Tare prompt, select Y(es) or N(o) to enable or disable auto clear
tare. Auto clear tare depends on the tare interlock condition.
If Tare Interlock and Auto Clear Tare Are Enabled
The terminal proceeds to the Display Tare prompt. Continue to step 8.
If Tare Interlock Is Disabled, and Auto Clear Tare Is Enabled
At the After Print? prompt, select Y(es) to clear tare after a print command is
issued, or select N(o) to clear tare at a predetermined threshold value.
If Auto Clear Tare After Print Is Disabled
1. Press ENTER at the Clear Threshold prompt; at the Wgt? prompt enter a unit value.
When the gross scale weight exceeds then falls below the threshold value, the
terminal automatically clears tare and returns to gross mode.
2. At the Check Motion? prompt, select Y(es) or N(o) to enable or disable the motion
check. If enabled, the terminal checks for stability of the load on the platform before
proceeding with auto clear tare.
3. At the Display Tare? prompt, select Y(es) or N(o) to enable or disable the tare
display. If enabled, the terminal displays the current tare value on the lower
display, provided the display is not showing other operator messages.
4. At the Net Sign Corr? prompt, select Y(es) or N(o) to enable or disable the net sign
correction feature.
5. At the JagIDNET Tare? prompt, select Y(es) or N(o) to enable or disable this
feature. NOTE: This prompt only occurs when you have an IDNET base. If you
If auto clear tare is enabled, the
terminal does not display the Clear
Threshold prompt.
Tare will not be displayed if an
auxiliar
y
unit is selected. Otherwise,
tare is displayed in the lower
display area.
**Applies only to IDNET
(MMR) bases.
Printing and tare operations will
w
ait until a stable condition exists
before proceeding with the action.
See Stability Detect sub-block in this
chapter for more information on
setting the sensitivity.
Tare will clear only at gross zero if
tare interlock and auto clear tare are
both enabled.