Technical data

METTLER TOLEDO JAGXTREME Terminal Technical Manual
(9/00) 3-22
If Auxiliary Units Are Enabled
NOTE: The alternate weight units selection is reset to factory default under the Scale
Interface Program Block, NOT the Application Environment Program Block.
1. At the Aux Unit? prompt, select the desired pre-programmed auxiliary unit
conversion factor or define a custom unit. The auxiliary unit selected here also
applies to the rate function. Selections include:
lb dwt
kg t
g ton
oz custom
lb-oz ozt
If Custom Unit Is Selected
1. At the Factor? prompt, enter a conversion factor. This factor is the number that will
by multiplied by the main units to calculate the custom unit. Some rounding error
may occur since this calculation uses a higher internal resolution to determine the
converted value. Make sure that the maximum converted value does not exceed
the display capacity of the lower display.
2. At the Name? prompt, enter the name for the custom unit, up to six characters.
3. At the Rate Enabled? prompt, select Y(es) or N(o) to enable or disable the rate
parameters for the custom unit. Rate parameters define the period of time
JAGXTREME terminal uses when displaying average weight change. If disabled,
JAGXTREME terminal continues to the next sub-block. If enabled, configure the next
two parameters.
4. At the Time Units? prompt, select the rate function time units. Options include:
Secaverage weight change per second
Minaverage weight change per minute
Houraverage weight change per hour
5. At the Int Period? prompt, use the numeric keys to enter the sampling period (in
seconds) that the terminal will use to calculate the average weight change for the
time units selected above. You must enter a value between 1 and 60.
6. At the Intrval? prompt, use the SELECT key to select 1/2 sec(ond),1 sec(ond), or 5
sec(onds). This determines how often the terminal calculates a new rate value.
Use 1 sec if you have a fast process which requires the most current rate value.
Use 5 sec when you want a smoother rate value.
7. Continue to the next sub-block or exit the setup mode.
Auxiliary Units must be enabled
to use the rate function.