Technical data
METTLER TOLEDO JAGXTREME Terminal Technical Manual
(9/00) 3-16
3. To exit this sub-block, press ENTER. To reset the parameters, press SELECT until
the display shows an N(o). Press ENTER. The display will then show RES ALL Y.
4. Press ENTER to confirm your choice. If you wish to exit without resetting the
parameters, press SELECT then ENTER.
5. If you have chosen to reset to factory defaults, POWER OFF appears in the display.
The JAGXTREME terminal must be powered down and then power reapplied.
6. The base will be calibrated internally, and the weighing platform will be assigned a
new scale number.
SCALE PARAMETERS
Program Sub-block
The display shows SCALE, allowing you to set certification capability, maximum
capacity, range, and graduation.
1. With Y(es) displayed, press ENTER.
2. The display shows NO APPR to allow selecting the scale for a non-certified mode
of operation. Use the SELECT key to change to APPR, then confirm your choice by
pressing ENTER.
3. The various types of K bases are displayed using the SELECT key. When the
appropriate base model number appears in the display, use the ENTER key to
select that choice.
4. The display shows CAP xxx kg, indicating the maximum capacity of the selected
weighing platform. The maximum capacity at which the scale operates is
selectable in three stages (full, half, or quarter). Depending on the country selected
in the NATION sub-block, the capacity may be displayed in pounds. Press SELECT
to step through the capacity choices. Press ENTER to choose the displayed
answer.
5. The readability of the base can be set using the SELECT key. When the display
shows MR, the display shows the display readability of the first range of a
multirange scale. When the display shows D, the displayed accuracy of a single
range base is shown.
LINEARITY Program Sub-
block
After a measuring cell is replaced, or the stored linearity code is changed, it must be re-
entered. The linearity code is located at the rear of the identcode bracket.
1. The display shows LINEA. Press ENTER to enter this sub-block.
2. The display indicates the current stored value LINEA +XXXX.
3. Choose the sign of the linearity code by using the SELECT key. When the proper
sign is displayed, use the ENTER key to confirm your choice.
4. Enter the linearity code using a combination of the SELECT key and the ENTER key.
The ENTER key will insert zeros in place of the dashes.
5. The SELECT key is used to increment the right-most zero to the desired number.
When the displayed value is correct, press ENTER.