Technical data

Chapter 3: Programming and Calibration
Scale Interface Program Block
(9/00) 3-7
Scale Interface Program
Block
The Scale Interface program block lets you set and calibrate the features that affect
weighing performance. The following diagram describes this block. Press ENTER at the
Scale Interface prompt to access the Scale Interface program block and configure the
sub-blocks.
*
Multiple Load Cell Scales Only.
**These menu selections are replaced by Servicemode and its associated menu when Scale
Type = High Precision.
1 Appears only if DigiTOL J-Box or POWERCELL selected in Scale Type sub-block.
Scale Interface
Market
Scale Type
Calibration Unit
Capacity
Increment Size
Select Market USA, EC,
AUS, CAN, NONE
NO. of Internal Scale
1, 2, 3, 4 or 5
Units pounds, kg,
grams, t
Capacity entry
Number of weighing
intervals
Scale Type
Analog, DigITOL, IDNET, D-J Box,
Ures-High, Ures-Low, Pwr Cell,
Summing, None
Location of Scale connection
Low increment
entry
Mid increment
entry
High increment
entry
!
!!
! !
!
!!
! ! !
**
**
**
**
**
**
**
**
**
**
**
Linearity
Correction
Calibration
Zero
Adjustment
Span
Adjustment
Gravity
Adjustment
Reset to
Factory
Enable Linearity
Y(es) or No
Empty the scale
Empty the scale
Enter test
weight value
Enter compensation
value
Are you sure?
Capture
Zero
A
dd test wei
g
ht enter
weight value
Capture
Span
Capture
Zero
Capture
Span
Add in Sum?
Y(es) or No
Shift Adjustment*
1
Empty Scale Capture Zero Load on Cell N Capture Cell