Technical data

Chapter 3: Programming and Calibration
General Information
(9/00) 3-3
If the Enter Setup? prompt is not displayed, try the following steps:
Remove AC power.
Remove the controller board and set S2-1 to the off position, or
Push in CAL switch while entering setup.
Replace the controller board.
Apply power to the terminal and repeat steps 1 and 2.
4. Once Scale Interface is displayed, press ENTER to open this block or press
SELECT to move to the next program block.
Navigating
To move through the various program blocks and sub-blocks:
1. Press SELECT to scroll through the program blocks. When the desired block is
displayed, press ENTER to open it.
2. Program blocks contain sub-blocks that handle specific areas of functionality. Once
you are in a program block, press the SELECT key to move through the sub-blocks.
Use the ENTER key to confirm your parameter option selection in the sub-blocks.
3. After configuring a sub-block, the terminal proceeds to the next sub-block. When
you finish configuring the last sub-block in a program block, the terminal proceeds
to the next program block. Continue through each program block until you have
finished configuring each one.
4. To return to normal operation after configuring the parameters in each sub-block,
press ESCAPE several times until the prompt Exit Setup? is displayed.
5. Press ENTER to confirm that you want to exit setup mode. If you do not wish to exit,
press SELECT to choose another program block within setup mode.
6. To protect the setup parameters from being changed once you have configured
them, power-down the terminal and install the W3 jumper on the Controller board.
You can exit your position within a program block by pressing ESCAPE at any time. You
may need to press ESCAPE several times to exit setup mode and continue to normal
operation.
Arrows in the upper display indicate your position within a program block. The following
chart describes the arrows relative to block position.
Arrows Block Position
1 Arrow τ You are in Setup Mode, top level.
2 Arrows ττ You are in a Program Block.
3 Arrows τττ You are in a sub-block.
4 Arrows ττττ You are configuring an element within a sub-block.
Note: Use anti-static protection
straps whenever touching
electronics boards.