Technical data
METTLER TOLEDO JAGXTREME Terminal Technical Manual
(9/00) 2-16
COM1 RS-232 (Controller PCB Serial Port)
The following diagram and table describe COM1 pin-to-pin cable connections using an
RS-232 cable. Maximum recommended cable length is 50 feet (15.24 meters).
JAGXTREME Terminal COM 1
!
TXD RS-232 Transmit
!
RXD RS-232 Receive
!
GND Signal Ground
!
CLTX+
!
CLRX+
!
CLRX-
Pin Connection for METTLER TOLEDO Devices Using COM1 RS-232
JAGXTREME
Terminal COM1
8622 8806 8842 8844 8855*** 8860**
8804** 8840 8843 8845 8856 8865 MP750
8617-TB2
9323-TB2
9325-TB2
8618
TXDA 3* 2 InputCom
RXDA
GND 7* 3 RS232 Input
CLTX+
CLRX+
CLRX−
*Each of these devices uses this connection.
**Pinout shown is for use without Plug In Adapter (8804 P/N 127358 00A, 8860 P/N 128019 00A).
***The 8855 using RS-232 must have the 129618 00A Interface PCB. The baud rate for the JAGXTREME terminal must be set to 300 baud. If
the interface PCB is part number 123654 00A or 137651 00A, the JAGXTREME terminal TXDA terminal must be connected to Pin 2 of the
8855 Interface PCB. In this case, set the JAGXTREME terminal baud rate to 1200.