Technical data
Chapter 2: Installation
Electrical Connections
(9/00) 2-9
The following diagrams describe DigiTOL terminal strip wiring.
Enhanced JAGXTREME Terminal COM2 or
COM4*Port
DigiTOL NC TXDB
J-Box
NC RXDB
TB2-12 Blue Gnd
TB1-2 Red TXD+
TB1-3 White TXD-
TB1-5 Black RXD+
TB1-4 Yellow RXD-
TB2-10 Green +20V
Chassis Ground
POWERCELL Connections (Non-Hazardous Area CMOS POWERCELL Applications)
Recommended maximum cable distance is 900 feet (274.32 meters) for all CMOS non-
hazardous applications regardless of the number of cells (assuming 16 or 20 gauge
wire).
POWERCELL JAGXTREME Terminal POWERCELL
J-Box
PCB
A COM A
B COM B
Gnd Gnd *
Gnd
Gnd
+VA +20 V*
+VB
+VC
For more details on installations in
hazardous areas using POWERCELL
Intrinsic Safety Barriers, refer to the
POWERCELL installation instructions
(P/N 142463 00A), and Print
TC100442 (included in the
POWERCELL installation instructions).
*There should be three +20V and
three ground wires in the cable
between the Junction Box and the
POWERCELL PCB.