Error Message Description Probable Cause Remedy
SER_IN_TIMEOUT There was a timeout waiting for
serial input.
The serial device talking to
the terminal has not sent the
required input to the terminal.
This could also be caused by
communication errors.
Verify message exchange
between the JAGXTREME
terminal and serial de vice.
SER_MSG_SEQ_ERR There was an error in the
sequencing of demand print
messages.
Most likely, when one
terminal is printing at a
remote terminal, a message
was lost in the Ethernet
communications.
If problem persists, check
Ethernet wiring, terminations,
and connections.
SERIAL_MSG_ERROR The Serial Services soft ware
modules got an in valid request.
JAGXTREME software error. Power down, then up. If error
still occurs, reset the
JAGXTREME terminal to factory
settings and reprogram setup
parameters.
SETPOINT_NO_RATE The user has configured a rate
setpoint, but has not configured
rate calculation.
The user has not configured
the rate calculation.
Configure the rate function in the
Alternate Weight Units sub-
block.
SHIFT_ADJUST_ERROR The shift adjustment factors could
not be calculated.
The weight placements
during the shift adjust
procedure was incorrect.
Carefully redo the shift
adjustment, perhaps, with
bigger weights.
SPX_NET_ERROR SPX returned a network error
status.
Most likely, there is a
problem with the Ethernet
wiring.
Check the network setup,
wiring, connections,
addressing, and terminations.
SYMMETRY OK SYMMETRY CHECK OK Previous symmetry error has
now gone back within the
established limits,
There may still be a potential
POWERCELL hardware problem.
You should perform diagnostics
checks the individual cells in
the symmetry set.
Syntax error JagBASIC programming error. The JagBASIC program has a
syntax error.
Correct the JagBASIC program.
TARE_ABOVE_LIMIT Tare value exceeds the allowed
limit.
In a legal for trade
environment in certain
markets, the tare value
cannot exceed the highest
weight in the lowest range of
a multi-range scale.
Check local legal-for-trade
requirements.
TARE_NOT_IN_INCR Keyboard Tare is not entered in a
rounded value to the nearest
increment.
In certain markets, the
keyboard tare value must be
entered in as a value
rounded to the nearest
increment.
Make sure the keyboard tare
value is rounded to the nearest
increment.
TARE_OVER_CAPCTY Tare exceeds the capacity of the
scale.
The tare value cannot exceed
the capacity of the scale.
Make sure the tare value is less